Comparison of different biomaterials for glaucoma drainage devices
R S Ayyala1, L E Harman, B Michelini-Norris
1Department of Ophthalmology, University of South Florida, Tampa, USA. rsayyala@hotmail.com
Archives of Ophthalmology (Chicago, Ill. : 1960)
|February 26, 1999
Summary
Silicone endplates caused the least inflammation in rabbit eyes compared to polypropylene and Vivathane. Biomaterial choice impacts inflammation and potential failure of glaucoma drainage devices.
Area of Science:
- Ophthalmology
- Biomaterials Science
- Glaucoma Research
Background:
- Glaucoma drainage devices (GDDs) are crucial for managing intraocular pressure.
- The inflammatory response to GDD components can influence surgical outcomes.
- Endplates within GDDs are in direct contact with ocular tissues, necessitating evaluation of their biocompatibility.
Purpose of the Study:
- To compare the inflammatory reactions elicited by silicone, polypropylene, and Vivathane endplates in a rabbit subconjunctival model.
- To assess the biocompatibility of a novel polymer, Vivathane, relative to established biomaterials.
Main Methods:
- Endplates of three biomaterials (silicone, polypropylene, Vivathane) were implanted into the rabbit subconjunctival space.
- Conjunctival hyperemia was graded masked to the treatment group.
- Histological and scanning electron microscopy analyses were performed after three weeks.
Main Results:
- Silicone demonstrated significantly less inflammation compared to polypropylene and Vivathane.
- Polypropylene and Vivathane groups exhibited higher clinical and histological inflammatory scores.
- Higher rates of extrusion were observed for polypropylene and Vivathane endplates.
Conclusions:
- Silicone is the most inert biomaterial among those tested for endplate applications in glaucoma surgery.
- Biomaterial-induced inflammation may be a contributing factor to glaucoma drainage device failure.
- Careful surgical handling of endplates is recommended to minimize iatrogenic tissue damage and inflammation.
Related Concept Videos
Glaucoma: Overview
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
Open Angle Glaucoma: Treatment
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...