Related Experiment Video
Updated: Aug 7, 2026

A Whole Cell Bioreporter Approach to Assess Transport and Bioavailability of Organic Contaminants in Water Unsaturated Systems
Published on: December 24, 2014
Improved method for the storage of groundwater samples containing volatile organic analytes
1ManTech Environmental Research Services Corporation, Environmental Protection Agency/National Risk Management Research Laboratory, P.O. Box 1198, Ada, Oklahoma 74821, USA.
Teflon septa in vials cause volatile organic analyte loss, impacting petroleum hydrocarbon testing accuracy. An alternative storage method using lead foil septa and a sodium phosphate preservative prevents analyte loss for at least 21 days.
Area of Science:
- Environmental Chemistry
- Analytical Chemistry
Background:
- Standard methods for analyzing volatile organic compounds (VOCs) in water often use Teflon septa in glass vials.
- Teflon's inertness is generally assumed, but its sorption properties for specific analytes require careful evaluation.
Purpose of the Study:
- To investigate the sorption of various volatile organic analytes by Teflon septa in standard sample vials.
- To assess the impact of this sorption on the accuracy of volatile petroleum hydrocarbon measurements.
- To develop and validate an alternative storage method to prevent analyte loss.
Main Methods:
- Analyzed volatile organic analytes including alkanes, aromatics, and chloroethenes in laboratory solutions and field samples.
- Evaluated analyte concentration changes in standard Teflon-lined vials stored at room temperature and 4°C over time.
- Tested an alternative storage protocol using lead foil septa and a sodium phosphate preservative.
Main Results:
- Significant loss (>10%) of n-alkanes and isoalkanes occurred within 24 hours at room temperature.
- Aliphatic analyte losses (>10%) were also observed over 21 days at 4°C.
- The alternative storage method successfully prevented analyte loss for at least 21 days at room temperature.
Conclusions:
- Sorption by Teflon septa can significantly reduce the accuracy of volatile petroleum hydrocarbon measurements in water samples.
- Traditional storage methods using Teflon septa are inadequate for preserving the integrity of certain volatile organic analytes.
- A novel storage protocol employing lead foil septa and a sodium phosphate preservative offers a reliable alternative for sample preservation.
More Related Videos
08:57VacuSIP, an Improved InEx Method for In Situ Measurement of Particulate and Dissolved Compounds Processed by Active Suspension Feeders
Published on: August 3, 2016
09:19Capturing Actively Produced Microbial Volatile Organic Compounds from Human-Associated Samples with Vacuum-Assisted Sorbent Extraction
Published on: June 1, 2022
Related Concept Videos
Volatilization
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...