Related Experiment Videos
Endoscopic threaded imaging port (EndoTIP) for laparoscopy: experience with different body weights.
1Department of Obstetrics and Gynecology, St. Joseph's Health Centre, Toronto, ON, Canada. artin@pathcom.com
Obesity Surgery
|March 5, 1999
Summary
A novel laparoscopic access cannula, requiring no trocar, demonstrated safety and efficacy in 234 patients. This trocar-less system is suitable for all body weights, including obese individuals, with no reported complications.
Area of Science:
- Minimally Invasive Surgery
- Surgical Access Techniques
- Laparoscopy
Background:
- A novel trocar-less laparoscopic access cannula was developed for primary port insertion.
- The system features a threaded cannula with a blunt tip, enabling radial fascial and muscle fiber spreading for insertion.
- Magnified visualization is provided during insertion and exit.
Purpose of the Study:
- To evaluate the safety and efficacy of a new trocar-less laparoscopic access cannula.
- To assess the device's performance across various patient weight categories, including obesity.
Main Methods:
- The cannula was utilized in 234 consecutive patients undergoing laparoscopic procedures.
- Patient cohorts included normal weight, mildly obese, moderately obese, and markedly obese individuals (BMI ≥35).
Main Results:
- No instrument-related or insertion-related complications occurred.
- Insertion was successful in all cases, with slightly longer times in morbidly obese patients with prior umbilical surgery.
- No port-site hernias were observed during 6-48 months of follow-up.
Conclusions:
- The reusable, trocar-less cannula is a safe and effective method for laparoscopic primary port insertion.
- The device is suitable for patients of all body weights, including those with obesity.