The QA pressure measurement system: an accuracy and reliability study.

D van Dijk1, G Aufdemkampe, S van Langeveld

  • 1Center Hospital Apeldoorn, The Netherlands.

Spinal Cord
|March 5, 1999
PubMed
Summary

The Queen Alexandra Pressure Measurement System (QA PMS) is accurate and reliable for measuring buttock pressures. Spinal cord injured (SCI) patients exhibit higher peak pressures during sitting compared to normal subjects.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Guiding complexity in V-A ECMO weaning decisions for cardiogenic shock: Insights from a conjoint analysis.

Perfusion·2026
Same author

[Autonomy promoting policy for complex chronic suicidality].

Tijdschrift voor psychiatrie·2024
Same author

Clinical decision making for VA ECMO weaning in patients with cardiogenic shock A formative qualitative study.

Perfusion·2024
Same author

Twelve years of circulatory extracorporeal life support at the University Medical Centre Utrecht.

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation·2021
Same author

Resting energy expenditure by indirect calorimetry versus the ventilator-VCO<sub>2</sub> derived method in critically ill patients: The DREAM-VCO<sub>2</sub> prospective comparative study.

Clinical nutrition ESPEN·2020
Same author

The effect of cisatracurium infusion on the energy expenditure of critically ill patients: an observational cohort study.

Critical care (London, England)·2020

Area of Science:

  • Biomedical Engineering
  • Rehabilitation Technology
  • Clinical Measurement

Background:

  • Accurate pressure measurement is crucial for preventing pressure ulcers, especially in individuals with spinal cord injuries (SCI).
  • The Queen Alexandra Pressure Measurement System (QA PMS) is a tool used for this purpose, but its performance requires validation.

Purpose of the Study:

  • To assess the accuracy and reliability of the QA PMS.
  • To compare buttock pressure measurements during sitting between SCI patients and normal subjects.

Main Methods:

  • The study involved an accuracy (calibration) and reliability (test-retest) assessment of the QA PMS.
  • 16 SCI patients and 15 normal subjects participated.
  • Statistical methods included Standard Error of the Mean (SEM), Technical Error of Measurement (TEM), Intraclass Correlation Coefficient (ICC), and t-tests.

Related Experiment Videos

Main Results:

  • The QA PMS demonstrated high accuracy (mean SEM = 0.30 mmHg) and good reliability (mean ICCs ranging from 0.85 to 0.92).
  • SCI patients showed significantly higher peak pressures (134.1 mmHg) compared to normal subjects (75.7 mmHg).
  • Normal subjects had higher mean pressures (31.0 mmHg) than SCI patients (28.5 mmHg).

Conclusions:

  • The QA PMS is a sufficiently accurate and reliable system for pressure measurement.
  • Significant differences in pressure distribution exist between SCI patients and normal subjects, with higher peak pressures in SCI patients being a key finding.