Biosensor for the enantioselective analysis of S-perindopril
Related Experiment Video
Updated: Oct 24, 2026

Use of Enzymatic Biosensors to Quantify Endogenous ATP or H2O2 in the Kidney
Published on: October 12, 2015
H Y Aboul-Enein1, R I Stefan, G L Radu
1Bioanalytical and Drug Development Laboratory, Biological and Medical Research Department, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia.
Abstract:
Because S-perindopril enantiomer is the eutomer which is responsible for the angiotensin-converting enzyme inhibition activity, it is necessary to develop a reliable method for its assay from its distomer, the R-enantiomer. For this purpose, an amperometric biosensor was developed based on L-amino acid oxidase. The working range of the described biosensor was 20pmol/L-10 micromol/L on the 7.0-7.4 pH range, with a detection limit of 2pmol/L. The low enantioselectivity for R-perindopril, as compared with S-enantiomer, was demonstrated by both mixed solutions and separate solutions methods (amperometric selectivity coefficient is 1.0 x 10(-4)). The biosensor was also selective towards D-proline and polyvinylpyrrolidone. The amperometric biosensor can be used for enantioselective analysis of S-perindopril in raw material, with an RSD < 1%. The life time (t95%) of the biosensor is three weeks.
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Antihypertensive Drugs: Angiotensin-Converting Enzyme Inhibitors