Prospective validation of single plasma sample 99mTc-ethylenedicysteine clearance in adults
L Kabasakal1, M Halaç, A F Yapar
1Department of Nuclear Medicine, Cerrahpaşa Medical Faculty, Istanbul, Turkey.
Unlabelled:
99mTc-L,L-ethylene, L, dicysteine (EC) clearance shows strong correlation with orthoiodohippurate clearance, and it is possible to estimate effective renal plasma flow from 99mTc-EC clearance. In routine clinical studies, it is practical to use the one or two plasma sample method instead of multiple plasma samples for clearance determination. A single-sample technique was developed for 99mTc-EC, and a regression formula was generated. A prospective study tested the validity of this regression formula.
Methods:
The study population was composed of 26 patients with a wide range of renal function. Multiple plasma sample 99mTc-EC clearances were calculated from all patients using the open two-compartment model. Single plasma sample clearances were also determined from the 54-min plasma sample using the regression formula published previously.
Results:
The multiple-sample plasma clearance of 99mTc-EC ranged from 46 to 668 mL/min with a mean of 300.76 +/- 150.73 mL/min. The clearances obtained from the 54-min plasma sample ranged from 49 to 699 mL/min, with a mean of 297.39 +/- 152.23 mL/min. There was an excellent correlation between the clearances obtained by the two techniques (r = 0.99, slope = 0.9911). The standard error of estimation was found to be 25.9 mL/min.
Conclusion:
This study suggests that 99mTc-EC clearance can be estimated from 54-min plasma samples with an acceptable error of estimation for most routine clinical studies.