Malfunction of Björk-Shiley valve prosthesis in tricuspid position
Abstract:
Eight months after triple valve replacement with Björk-Shiley tilting disc valves a patient developed symptoms and signs suggesting malfunction of the prosthesis in the tricuspid position. This was confirmed by echocardiography and angiocardiography, and at operation thedisc of the prosthesis was found to be stuck half-open by fibrin and clot. A further 11 patients with the same tupe of prosthesis in the triscupid position were then studied by phonocardiography and echocardiography. In one of these the prosthesis was found to be stuck and this was confirmed by angiocardiography and surgery. These 2 cases are reported in detail and the findings in the other 10 are discussed. The implications of this high incidence of malfunction of the Björk-Shiley prosthesis in the tricuspid position are considered. Echocardiography appears to be essential in the follow-up of such patients.
More Related Videos
08:42Cox-Maze IV Procedure Concomitant with Valvular Surgery In Situs Inversus Dextrocardia: A Single-Center Experience in China
Published on: February 11, 2022
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Mitral Regurgitation III: Medical Management
Mitral Stenosis III: Medical Management