Corneal indentation during ultrasonic pachometry.

O D Solomon1

  • 1Department of Ophthalmology, Case Western Reserve School of Medicine, Cleveland, Ohio, USA. info@eyeberspace.com

Cornea
|March 25, 1999
PubMed
Summary

Ultrasonic pachometry for measuring corneal thickness is not affected by probe indentation. This finding is crucial for accurate ophthalmic diagnostics and intraocular pressure assessment.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Corneal swelling and extended wear.

The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc·1999
Same author

A 3-year prospective study of the clinical performance of daily disposable contact lenses compared with frequent replacement and conventional daily wear contact lenses.

The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc·1996
Same author

Corneal stress test for extended wear.

The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc·1996
Same author

Extended wear, corneal hypoxia, and corneal ulcers.

The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc·1994
Same author

Testing hypotheses for risk factors for contact lens-associated infectious keratitis in an animal model.

The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc·1994
Same author

Micropachometry.

The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc·1993
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Area of Science:

  • Ophthalmology
  • Biomedical Engineering
  • Medical Instrumentation

Background:

  • Accurate corneal thickness measurement is vital for ophthalmic diagnostics.
  • Corneal indentation during ultrasonic pachometry is a potential source of measurement error.
  • Understanding the impact of indentation is essential for reliable pachymetry.

Purpose of the Study:

  • To investigate the effect of corneal indentation by an ultrasonic probe on corneal thickness measurements.
  • To determine if varying indentation forces influence ultrasonic pachometry readings.

Main Methods:

  • An ultrasonic transducer was integrated with a Goldmann Applanation Tonometer.
  • Corneal thickness was measured in five human subjects at different tonometer tension settings (0, 10, 20, 40, 80).
  • Radio-frequency signals were captured and thicknesses digitally measured.

Main Results:

  • No significant trend of decreasing or increasing corneal thickness measurements was observed with increasing tonometer tension.
  • Corneal indentation did not introduce a systematic bias in the measurements.

Conclusions:

  • Corneal thickness measurements obtained via ultrasonic pachometry are not significantly affected by the indentation forces of the ultrasonic probe.
  • The findings support the reliability of ultrasonic pachometry for corneal assessment, even with probe contact.