Primary total hip arthroplasty with the AML total hip prosthesis
Clinical Orthopaedics and Related Research
|April 2, 1999
Summary
This study on uncemented total hip arthroplasty found a 15% revision rate, with significant polyethylene wear and osteolysis observed in many cases after an average of 11.3 years. Hip function improved, but component longevity remains a concern.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- Total hip arthroplasty (THA) is a common procedure for hip joint pathology.
- Uncemented prostheses aim for biological fixation, potentially offering long-term benefits.
- Assessing the long-term performance of specific uncemented hip implant designs is crucial for clinical decision-making.
Purpose of the Study:
- To evaluate the clinical outcomes and implant survival of the AML total hip prosthesis in primary THA.
- To investigate the incidence of osteolysis, polyethylene wear, and revision rates associated with this specific uncemented implant system.
- To correlate radiographic findings with functional outcomes in patients undergoing THA with the AML prosthesis.
Main Methods:
- A prospective study involving 50 patients (52 hips) undergoing primary THA with an uncemented AML prosthesis (trispiked cup, porous coated stem).
- Patients were followed for an average of 11.3 years, with clinical assessment using the Harris Hip Score.
- Radiographic evaluation included assessment for osteolysis, component migration, and polyethylene wear.
Main Results:
- The average Harris Hip Score improved significantly from 59 preoperatively to 90 postoperatively.
- Overall revision rate was 15% (eight hips), primarily due to acetabular component issues (15%).
- Significant findings included acetabular and femoral osteolysis in 38% of hips and polyethylene wear averaging 0.29 mm/year.
Conclusions:
- While the AML uncemented total hip prosthesis demonstrated good functional improvement, a notable rate of revision and concerning incidence of osteolysis and polyethylene wear were observed.
- The high rate of acetabular component revision suggests potential challenges with fixation or wear in this specific design.
- Long-term surveillance is essential to monitor the durability and identify potential failure modes of uncemented hip prostheses, particularly regarding polyethylene wear and osteolysis.