Organic buildup and residual blood on infant stethoscopes in maternal-infant areas
1School of Nursing at Loma Linda University, CA 92350, USA.
Objective:
To identify the presence of residual blood and organic matter on "clean" stethoscopes in maternal-infant units.
Design:
In this retrospective, nonexperimental study, stethoscopes were tested using qualitative measurements.
Setting:
Using a nonprobability sampling technique, 11 acute care hospitals in a three-state area of the southwestern United States were studied.
Participants:
All stethoscopes found on the maternal-infant units were included, for a total sample size of 97.
Instruments:
A hand-held 10-power lens was used to visually rank the amount of organic buildup, and the phenolphthalein test was used to detect residual blood on the stethoscope.
Results:
Of 97 clean infant stethoscopes, 80% of labor and delivery and 72% of nursery stethoscopes had organic buildup on the diaphragm. Both areas had similar rates of organic buildup, chi2 (1, N = 97) = 1.00, p = ns. Nursery areas did have significantly lower rates of residual blood than stethoscopes from labor and delivery, phi2 (1, N = 97) = 9.89, p = .002. Seventy-six percent of labor and delivery stethoscopes were positive for blood, as compared to 46% of nursery stethoscopes.
Conclusions:
Traditional methods for cleaning stethoscopes used in labor and delivery and nursery areas are ineffective in removing blood and other body fluids from the stethoscope.