Contamination of the medical air supply with oxygen: a clinical engineering incident investigation
W E Kane1, J D Ridley, M W Sheehan
1Biomedical Engineering Department, Royal Jubilee Hospital, Victoria, British Columbia.
Journal of Clinical Engineering
|July 1, 1990
Abstract:
Two incidents of oxygen contamination in the hospital medical air supply occurred on consecutive weekends. This paper describes those incidents and the investigation that followed. The cause of the contamination was determined to be faulty check valves in some Bird air/oxygen blenders. The investigation of the incidents was hindered because the conventional blender test methods did not detect the problem. Only after monitoring the medical air and oxygen pressures and simulating the real conditions did the cause of the problem surface.