Evaluation of ten disposable manual resuscitators
1College of Pharmacy and Allied Health Professions, Northeastern University.
Respiratory Care
|October 1, 1990
Summary
This study evaluated 10 disposable resuscitators against ASTM Standard F-920. Only four models met all performance and safety requirements for reliable use.
Area of Science:
- Emergency Medicine
- Medical Device Performance
- Respiratory Care
Background:
- Disposable resuscitators are crucial for emergency ventilation.
- Evaluating their performance against established standards is vital for patient safety.
- Previous assessments of disposable resuscitator compliance with ASTM F-920 are limited.
Purpose of the Study:
- To assess the performance and safety of ten disposable resuscitators (six adult, four pediatric) against ASTM Standard F-920.
- To identify which disposable resuscitators meet the stringent requirements for use as replacements for permanent devices.
Main Methods:
- Ten disposable resuscitators were tested: SPUR, Code Blue, 1st Response, Hospitak MPR, CPR Bag, Pulmanex (adult); CPR Bag, 1st Response, Hospitak MPR, LSP Bag Mask (pediatric).
- Testing followed the American Society for Testing and Materials (ASTM) Standard F-920, utilizing a Bio-Tek VT-1 Ventilator Tester lung model.
- Evaluations included ventilation parameters (VT, f, I:E), fractional delivered oxygen concentration (FDO2) with and without attachments, valve function after contamination, backward leakage, pressure-limit, and mechanical shock resistance.
Main Results:
- All resuscitators met basic ventilation requirements (VT, f, I:E) and FDO2 with attachments.
- Nine out of ten met FDO2 standards without attachments.
- Three pediatric models failed the pressure-limit test, and four models (adult and pediatric) failed the mechanical shock test.
Conclusions:
- Only Code Blue, 1st Response, Pulmanex (with tube-type reservoir), and SPUR met all ASTM Standard F-920 requirements.
- These four models are deemed acceptable replacements for permanent resuscitators.
- Several disposable resuscitators failed critical safety and performance benchmarks, highlighting the need for careful selection.