Related Experiment Videos
Beyond the basics: operating the new generation of ventilator. A look at the features and functions of these units
R C Bone1, D H Eubanks, E Gluck
1Rush Medical College, Chicago.
Abstract:
Most modern ventilators have several key features in common: microprocessor control of operational and monitoring functions; electromechanical valves to control and adjust gas flow patterns; and extensive monitoring systems. In addition, these machines can provide a number of different modes of ventilation (including pressure support). Though not microprocessor-controlled, the Siemens Servo 900 series ventilators use feedback electronics to adjust inspiratory flow based on expiratory flow to meet preset volumes. In contrast, the Bennett 7200 units use microprocessor-regulated solenoid valves to deliver preset tidal volume. High-frequency ventilators deliver smaller tidal volumes at rates greater than 60 bpm.