Techniques for IABP timing, use--and discontinuance. Counterpulsation can reduce ischemia and improve hemodynamics
1University of Chicago Pritzker School of Medicine.
Abstract:
By restoring the balance between myocardial oxygen supply and demand, intra-aortic balloon counterpulsation can aid the ischemic or failing left ventricle. Rapid inflation and deflation of the balloon reduces afterload and improves coronary perfusion. Timing of inflation and deflation is adjusted according to the arterial pressure waveform. The balloon should inflate just after the aortic valve closes and deflate just before the aortic valve opens; early or late balloon inflation or deflation blunts the therapeutic effects. Discontinue the IABP support gradually; monitor the patient frequently for any signs of hemodynamic compromise, redevelopment of ischemic chest pain, or electrocardiographic changes.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...