Related Experiment Video
Updated: Aug 2, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
New whole blood methods and instruments: glucose measurement and test menus for critical care
Abstract:
A glucose oxidase-H 2O 2 substrate-selective electrode (SSE) along with eight other electrodes allows the Stat Profile 5 (SP5) to measure glucose, electrolytes, blood gases, pH, and hematocrit simultaneously in whole blood. For heparinized blood gas syringe samples (n=178), orthogonal (Deming) regression showed y=0.43 + 1.00x (mmol/L), where y is whole blood glucose, and x is plasma glucose measured with the Ektachem 400 glucose oxidase slide method. For heparinized plasma (n=197), y=0.25 + 1.02x. Least squares linear regression showed y=0.25 + 1.00x (s yxx=0.49) for whole blood versus plasma, y = 0.30 + 1.01x (s yxx=0.49) for plasma versus plasma, and r=0.994 for both cases. Whole blood and plasma glucose in the same samples measured by SSE differed by less than 1%. The SP5 combines the largest number of whole blood tests currently available in a rapid response instrument well-suited for care of the critically ill patient.
More Related Videos
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Hypoglycemia and Glucagon
Hyperglycemia