Investigation of fatigue failure of S-65-HL "Super Tygon" roller pump tubing
E J Snyder1, D L McElwee, H M Harb
1Thomas Jefferson University Hospital, Department of Biomedical Instrumentation, Philadelphia, PA 19107, USA.
Abstract:
A failure analysis was performed on Norton S-65-HL Tygon tubing. Fatigue testing was performed on four sizes of this tubing, and essentially showed how the tubing wears out. A dynamic "life hours to failure" test, which was performed on the 3/8" internal diameter (ID) size, quantified when the tubing ruptured. Based on results of laboratory testing and the institution's clinical extracorporeal membrane oxygenation (ECMO) experience, a reasonable life expectancy for the 3/8" S-65-HL Tygon size was determined for use in this institution's neonatal ECMO system. An understanding of the expected performance of roller pump tubing-an integral component of the ECMO system-is imperative to providing safe, effective extracorporeal life support.
More Related Videos
Related Concept Videos
Fatigue
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress Concentrations in Circular Shafts
Thin-Walled Hollow Shafts
Design Example: Deciding Thickness of Lubricating Fluid in a Shaft
To calculate the required thickness of the lubricant layer, the tangential velocity at the shaft's surface must first be determined. This velocity is calculated by converting the rotational speed to angular velocity...
Plastic Deformation in Circular Shafts