Related Experiment Video
Updated: Aug 15, 2026

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Passive surveillance of heart valve devices: Björk-Shiley outlet strut fracture rates
G L Grunkemeier1, W N Anderson
1Providence Health System, Albert Starr Academic Center for Cardiac Surgery, Portland, Oregon 97225, USA.
Abstract:
The Björk-Shiley Convexo-Concave heart valve has a mechanical failure mode called outlet strut fracture (OSF). Previous analyses have found that the hazard function for OSF is constant over time. This article uses the valve manufacturers implant database to examine the shape of the hazard function over time, employing a previously developed model to impute patient survival times. Decisions about patient management (e.g., elective explant) are based on projecting this risk into the future. An analysis that incorporates the decrement of patients at risk due to death and explant provides estimates of the actual occurrences of OSF over calendar time.

