New roller pump disposable provides safety and simplifies occlusion setting.
K Lee-Sensiba1, N Azzaretto, C Carolina
1North Shore University Hospital-New York University School of Medicine, Manhasset 11030, USA.
The Journal of Extra-Corporeal Technology
|February 6, 1997
Summary
The Better-Header, a novel arterial roller pump insert, enhances patient safety by automatically limiting outlet pressure and simplifying occlusion setting. This disposable device ensures predictable performance and alerts users to potential overpressurization events.
Area of Science:
- Biomedical Engineering
- Cardiovascular Technology
Background:
- Standard arterial roller pump tubing lacks integrated safety features for pressure regulation.
- Setting pump occlusion can be inconsistent, potentially leading to non-optimal blood flow dynamics.
Purpose of the Study:
- To introduce and evaluate the Better-Header, a disposable insert designed to enhance safety and functionality of arterial roller pumps.
- To assess the Better-Header's ability to limit outlet pressure and facilitate consistent occlusion settings.
Main Methods:
- The Better-Header, incorporating a Starling-like pressure relief valve, was integrated with standard arterial header tubing.
- Laboratory evaluation of pressure-flow characteristics was performed, including testing under clamped arterial line conditions.
- Clinical performance was assessed in over 500 cases at North Shore University Hospital.
Main Results:
- The Better-Header effectively limited arterial line pressure to safe, user-settable levels, even under high flow or clamped conditions.
- The device demonstrated consistent nonocclusive settings with predictable retrograde flow.
- Clinical use successfully alerted staff to high-pressure situations via fluid shunting and audible alarms.
Conclusions:
- The Better-Header offers significant safety improvements over standard arterial roller pump setups by providing automatic pressure limitation.
- It enables reliable and consistent nonocclusive settings, contributing to improved patient safety and procedural efficiency.
- The device has proven effective and reliable in extensive clinical application.