Single-dose breath-hold gadolinium-enhanced three-dimensional MR angiography of the renal arteries
V S Lee1, N M Rofsky, G A Krinsky
1Department of Radiology, New York University Medical Center, New York, NY 10016, USA.
Purpose:
To evaluate the quality of single-dose breath-hold three-dimensional (3D) magnetic resonance (MR) angiography of the renal arteries optimized with a 1-mL test bolus timing examination.
Materials And Methods:
Three-dimensional spoiled gradient-echo imaging (3.8-4.2/1.3-1.7 [repetition time msec/echo time msec], 25 degrees-40 degrees flip angle) was performed in 60 patients after administration of gadopentetate dimeglumine (average dose, 0.11 mmol/kg). Synchronization of contrast material administration with data acquisition was achieved with a 1-mL test dose of contrast material to estimate patient circulation parameters. Image quality was assessed by using contrast-to-noise (CNR), relative vascular enhancement, and venous-to-arterial enhancement ratios and subjective scoring of arterial and venous enhancement. The effect of the contrast material injection rate and the influence of breath holding during the timing examination also were examined.
Results:
Overall, of 60 studies, 58 were diagnostic and 56 demonstrated excellent arterial enhancement. Venous enhancement was seen in eight studies. The average aortic relative vascular enhancement (+/- SD) was 14.6 +/- 5.9, with an aorta-to-inferior vena cava (IVC) CNR of 69.7 +/- 43.9. The IVC-to-aorta venous-to-arterial enhancement ratio averaged 0.08 +/- 0.16. There was no significant difference in image quality based on injection rates or the performance of breath holding during the timing examination (P > .1).
Conclusion:
Breath-hold gadolinium-enhanced renal MR angiography free of venous enhancement can be performed consistently and reliably with 20 mL of contrast material when studies are synchronized to patient circulation time by using a timing examination.
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT