Effect of breathing circuit resistance on the measurement of ventilatory function
Related Experiment Video
Updated: Oct 23, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
D P Johns1, C M Ingram, S Khov
1Department of Respiratory Medicine, Alfred Hospital, Melbourne, Victoria, Australia.
Background:
The American Thoracic Society (ATS) has set the acceptable resistance for spirometers at less than 1.5 cm H2O/l/s over the flow range 0-14 l/s and for monitoring devices at less than 2.5 cm H2O/l/s (0-14 l/s). The aims of this study were to determine the resistance characteristics of commonly used spirometers and monitoring devices and the effect of resistance on ventilatory function.
Methods:
The resistance of five spirometers (Vitalograph wedge bellows, Morgan rolling seal, Stead Wells water sealed, Fleisch pneumotachograph, Lilly pneumotachograph) and three monitoring devices (Spiro 1, Ferraris, mini-Wright) was measured from the back pressure developed over a range of known flows (1.6-13.1 l/s). Peak expiratory flow (PEF), forced expiratory flow in one second (FEV1), forced vital capacity (FVC), and mid forced expiratory flow (FEF25-75%) were measured on six subjects with normal lung function and 13 subjects with respiratory disorders using a pneumotachograph. Ventilatory function was then repeated with four different sized resistors (approximately 1-11 cmH2O/l/s) inserted between the mouthpiece and pneumotachograph.
Results:
All five diagnostic spirometers and two of the three monitoring devices passed the ATS upper limit for resistance. PEF, FEV1 and FVC showed significant (p < 0.05) inverse correlations with added resistance with no significant difference between the normal and patient groups. At a resistance of 1.5 cm H2O/l/s the mean percentage falls (95% confidence interval) were: PEF 6.9% (5.4 to 8.3); FEV1 1.9% (1.0 to 2.8), and FVC 1.5% (0.8 to 2.3).
Conclusions:
The ATS resistance specification for diagnostic spirometers appears to be appropriate. However, the specification for monitoring devices may be too conservative. PEF was found to be the most sensitive index to added resistance.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Respiratory Volumes and Capacities