Continuous thermodilution cardiac output: clinical validation against a reference technique of known accuracy
E R Schmid1, D Schmidlin, M Tornic
1Division of Cardiovascular Anaesthesia, Institute of Anaesthesiology, University Hospital of Zurich, Switzerland. edith.schmid@ifa.usz.ch
Objective:
To evaluate the accuracy and precision of continuous thermodilution (CCO) by using a validated bolus thermodilution (BCO) reference technique as criterion standard.
Design:
Under circulatory steady state conditions, a CCO system (Vigilance, software versions 4.35 and 4.39) was validated with regard to CCO as well as iced and room temperature BCO.
Setting:
Intensive care unit at a university hospital.
Patients:
Method comparison was conducted in 56 cardiac surgical patients, 28 patients being allocated to one of the two software versions, and 14 within each group to either iced or room temperature BCO.
Measurements And Results:
CCO readings were registered in duplicate before and after three to five bolus injections conducted with both the Vigilance and reference systems. Iced BCO showed excellent agreement between the Vigilance and reference systems, yielding SDs of bias of 0.41 and 0.37 l/min and linear correlation coefficients (r) of 0.97 and 0.96. With room temperature BCO, agreement was significantly less. CCO, irrespective of software version, showed higher SDs of bias (0.90 and 0.84 l/min) and lower r values (0.84 and 0.81) than iced BCO (p < 0.0001). CCO measurements with software version 4.39 yielded a similar SD of bias to that with room temperature BCO.
Conclusion:
Decreased precision of CCO as compared to iced BCO may, in clinical settings, be outweighed by the advantages of automated and continuous monitoring. Under research conditions, however, iced BCO remains the method of choice.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
06:10Author Spotlight: Assessment of Cardiac Output Calculation by Thermodilution in Pigs for Effective Perfusion Flow During EVLP
Published on: June 28, 2024
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Catheterization II: Right Heart Catheterization