Determination of a calibration factor for the nondestructive assay of Guidant 32P brachytherapy sources
R Collé1, B E Zimmerman, C G Soares
1Physics Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899-8462, USA. rcolle@nist.gov
Abstract:
A calibration factor ('dial setting') for the nondestructive assay of Guidant TiNi-encapsulated 32P intravascular brachytherapy wire sources has been determined for measurements with the Capintec CRC-12 (sic. 'dose calibrator') ionization chamber. The calibration factor was derived from ionization current measurements with the CRC-12 followed by very quantitative, destructive assays of the 32P content in two sources.