Does ductal occlusion with the gianturco coil cause left pulmonary artery and/or descending aorta obstruction?
D Stromberg1, R Pignatelli, G L Rosenthal
1Department of Pediatrics, Baylor College of Medicine and Texas Children's Hospital, Houston, USA.
Insights
Gianturco coils effectively occlude patent ductus arteriosus. Despite 2D echocardiography showing coil impingement, Doppler and catheterization confirm no significant obstruction in pulmonary artery or aorta.
Area of Science:
- Pediatric Cardiology
- Interventional Cardiology
- Medical Imaging
Background:
- Patent ductus arteriosus (PDA) is a common congenital heart defect.
- Transcatheter coil occlusion is a minimally invasive treatment for PDA.
Purpose of the Study:
- To evaluate the safety and efficacy of transcatheter Gianturco coil occlusion for PDA.
- To assess potential vascular obstruction using echocardiography and catheterization data.
Main Methods:
- Thirty-two pediatric patients underwent transcatheter Gianturco coil occlusion for PDA.
- Transthoracic echocardiography (2D and pulsed Doppler) was performed post-procedure and at follow-up.
- Echocardiographic findings were compared with angiographic and hemodynamic data.
Main Results:
- 2D echocardiography frequently showed coil protrusion into the left pulmonary artery (90%) and descending aorta (59%).
- Pulsed Doppler and catheterization data revealed no significant obstruction to flow in these vessels (97% and 96% normal flow, respectively).
- No significant difference in flow velocities was observed between patients with and without apparent coil impingement.
Conclusions:
- Transcatheter Gianturco coil occlusion is safe and effective for PDA.
- 2D echocardiography may overestimate vascular obstruction; pulsed Doppler and catheterization are crucial for accurate assessment of flow dynamics.
Abstract:
Thirty-two patients (median age 4.5 years) underwent transcatheter Gianturco coil occlusion of a patent ductus arteriosus. Transthoracic echocardiography was performed the day after coil placement and at intermediate follow-up (median 8.6 months). Echocardiographic results were compared with angiographic and hemodynamic data obtained during catheterization. Two-dimensional (2D) echocardiography performed the day after ductal occlusion displayed evidence of coil protrusion into the left pulmonary artery in 28 of 31 patients (90%) and into the descending aorta in 17 of 29 (59%). However, pulsed Doppler analysis demonstrated normal left pulmonary arterial flow velocities in 28 of 29 patients (97%) and normal descending aortic flow velocities in 26 of 27 (96%). Pulse Doppler results were corroborated by angiographic and hemodynamic catheterization data, which showed no evidence of adjacent vessel obstruction. Peak Doppler velocities among patients with and without 2D echocardiographic left pulmonary artery or descending aorta coil impingement did not differ significantly. The discrepancy between 2D and pulse Doppler findings did not change significantly at intermediate follow-up. Thus, transcatheter occlusion of the patent ductus arteriosus with properly implanted Gianturco coils does not cause significant obstruction to flow in the left pulmonary artery or descending aorta despite frequently misleading 2D echocardiographic images of coil impingement on these vessels.