Increasing observer objectivity with audio-visual technology: the Sphygmocorder
Abstract:
The most fallible component of blood pressure measurement is the human observer. The traditional technique of measuring blood pressure does not allow the result of the measurement to be checked by independent observers, thereby leaving the method open to bias. In the Sphygmocorder, several components used to measure blood pressure have been combined innovatively with audio-visual recording technology to produce a system consisting of a mercury sphygmomanometer, an occluding cuff, an automatic inflation-deflation source, a stethoscope, a microphone capable of detecting Korotkoff sounds, a camcorder and a display screen. The accuracy of the Sphygmocorder against the trained human observer has been confirmed previously using the protocol of the British Hypertension Society and in this article the updated system incorporating a number of innovations is described.
More Related Videos
06:12Recording Horizontal Saccade Performances Accurately in Neurological Patients Using Electro-oculogram
Published on: March 13, 2018
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.