Effect of different AV delays on left ventricular diastolic function and ANF levels in DDD paced hypertensive

F K Panou1, N V Kafkas, N A Michelakakis

  • 1Department of Cardiology, General Hospital of Athens G Gennimatas, Greece.

Insights

Short atrioventricular delays during DDD pacing improve left ventricular diastolic function in hypertensive patients with heart block. This is shown by hormonal and echocardiographic measurements, with higher ANF levels during longer delays.

Area of Science:

  • Cardiology
  • Electrophysiology
  • Diabetology

Background:

  • Hypertension and left ventricular hypertrophy (LVH) are common.
  • Complete heart block necessitates pacemaker implantation.
  • Atrioventricular (AV) delay optimization is crucial for cardiac function in paced patients.

Purpose of the Study:

  • To investigate the impact of different atrioventricular (AV) delays during DDD pacing on left ventricular (LV) diastolic function.
  • To assess changes in atrial natriuretic factor (ANF) plasma levels and echocardiographic parameters with varying AV delays.

Main Methods:

  • Studied 15 hypertensive patients with LVH and complete heart block, paced in DDD mode.
  • Evaluated two-dimensional and Doppler echocardiography at 100 ms and 200 ms AV delays.
  • Measured ANF plasma levels at rest and peak exercise during different AV delay settings.

Main Results:

  • Higher ANF plasma levels observed at rest and exercise with longer (200 ms) AV delays (P < 0.001 and P < 0.0001, respectively).
  • ANF levels significantly increased with exercise compared to rest for both AV delays.
  • Short (100 ms) AV delays improved diastolic function, indicated by decreased isovolumic relaxation time (P < 0.001) and E/A ratio (P < 0.05).

Conclusions:

  • Short AV delays (100 ms) during DDD pacing appear to improve left ventricular diastolic function.
  • Hormonal (ANF) and echocardiographic measures support improved diastolic function with shorter AV delays.
  • Patients with smaller left ventricles showed a greater ANF increase with longer AV delays.