Problems with telemetry monitoring systems.
Journal of Clinical Engineering
|September 9, 1978
Summary
Telemetry monitoring systems in operating rooms experience frequent failures, averaging one every three days. These issues stem from battery depletion, lead problems, and mechanical failures, making telemetry unreliable for surgical environments.
Area of Science:
- Biomedical Engineering
- Clinical Engineering
- Medical Device Technology
Background:
- Purchasers of operating room (OR) and recovery room telemetry monitoring systems lack practical usage and ownership information.
- Telemetry systems are crucial for patient monitoring during surgical procedures and recovery.
Purpose of the Study:
- To investigate the practical aspects of ownership and usage of telemetry monitoring systems in OR and recovery settings.
- To identify common failure modes and their causes in OR telemetry systems.
Main Methods:
- Recorded 76 telemetry failures over six months across 18 telemetry channels in OR and recovery rooms.
- Analyzed failure causes, including operator error, machine malfunction, lead/electrode issues, battery depletion, and component failures.
- Documented repair data for transmitters and receivers over a two-year period.
Main Results:
- An average of one telemetry failure occurred every three days or 60 surgical procedures.
- Identified key failure contributors: lead/electrode problems (28%), battery depletion (25%), component failures (22%), and control setting/frequency issues (12%).
- Observed hazards include transmitter immersion in liquids, potential for displaying incorrect patient ECG data, and frequent transmitter loss.
Conclusions:
- Telemetry systems exhibit poor reliability and high costs compared to hardwired alternatives in the OR.
- Waterproofing, lead-failure warning circuitry, and improved transmitter design are recommended for OR telemetry.
- The risk of displaying incorrect patient data necessitates careful use and supplementary monitoring methods.