Related Experiment Videos
[Automatic regulator of venous pressure and venous outflow in the perfusion system]
Abstract:
A scheme for automatic regulation of the venous pressure and venous blood outflow during extracorporeal circulation is proposed. The system consists of a photoelectric sensor placed on a tube led out of the major venous trunkline, a converter and an electromechanical eccentric clamp that compresses the venous trunkline, all of which secures stabilization of the controlled values.