Intravenous nitroglycerin delivery: dynamics and cost considerations.
Hospital Pharmacy
|March 11, 1985
Summary
Polyvinylchloride (PVC) intravenous (IV) sets significantly reduce nitroglycerin (NTG) availability. Polyethylene IV sets ensure accurate NTG delivery, proving more effective and cost-efficient for infusions.
Area of Science:
- Pharmacology and Pharmaceutics
- Intravenous Drug Delivery Systems
Background:
- Nitroglycerin (NTG) is a critical vasodilator used in various clinical settings.
- The material of intravenous (IV) administration sets can affect drug stability and delivery.
- Previous studies suggest potential adsorption of drugs to polyvinylchloride (PVC) IV sets.
Purpose of the Study:
- To evaluate the impact of PVC versus polyethylene IV administration sets on nitroglycerin (NTG) delivery accuracy.
- To determine the relative availability of NTG when infused through different IV set materials at varying rates.
- To assess the cost-effectiveness of using polyethylene IV sets for NTG infusions.
Main Methods:
- Nitroglycerin solutions were infused through PVC and polyethylene IV sets at 6, 12, and 24 ml/hour.
- Delivered NTG concentrations were quantified using high-performance liquid chromatography over 24-hour infusions.
- Relative NTG availability was calculated by comparing delivered to initial concentrations.
Main Results:
- Polyethylene IV sets demonstrated high NTG relative availability (96.3–97.3%) across all infusion rates.
- PVC IV sets showed significantly lower NTG relative availability (41.5–76.0%), varying with infusion rate.
- NTG concentration delivered through PVC sets varied considerably over time and with infusion rate.
Conclusions:
- Polyethylene IV administration sets provide accurate and consistent delivery of nitroglycerin.
- PVC IV sets lead to substantial under-delivery of nitroglycerin, impacting therapeutic efficacy.
- Polyethylene IV sets are recommended for nitroglycerin infusions to ensure accurate dosing and cost-effectiveness.