Bacterial output from three respiratory therapy humidifying devices
Respiratory Care
|February 11, 1979
Summary
Hospital humidifiers were tested for bacterial contamination. Evaporative humidifiers showed no viable organisms, while nebulizers produced heavily contaminated bacterial aerosols, posing a significant risk to patients.
Area of Science:
- Medical Microbiology
- Infectious Disease Control
- Respiratory Therapy
Background:
- Hospital-acquired infections are a major concern.
- Air humidification devices are essential for patient care but can harbor microorganisms.
- Evaluating the microbial output of these devices is critical for patient safety.
Purpose of the Study:
- To assess the relative hazard of different air humidification devices used in hospitals.
- To quantify bacterial output from evaporative humidifiers versus nebulizers.
- To determine the risk of bacterial aerosolization from these devices.
Main Methods:
- Three common air humidification devices were evaluated: two evaporative types and one nebulizer.
- Device reservoirs were inoculated with Pseudomonas cepacia.
- Bacterial output in the gas stream was sampled using an Andersen six-stage bacterial sampler.
- Sampling occurred after a warm-up period for the nebulizer.
Main Results:
- Nebulizers produced heavily contaminated bacterial aerosols (>1000 P. cepacia colonies) even after warming.
- Evaporative humidifiers yielded virtually no viable organisms.
- Bacterial output was significantly higher from the nebulizer compared to evaporative devices.
Conclusions:
- Nebulizers pose a significant risk of delivering contaminated bacterial aerosols to hospitalized patients.
- Evaporative humidifiers are a safer alternative regarding bacterial output.
- Infection control protocols should consider the type of humidification device used.