Reliability of three different brands of adult volume ventilators in a one-year period
Abstract:
The maintenance records covering a 12-month period for 16 Puritan-Bennett MA-1, 7 Bourns BEAR-1, and 3 Siemens-Elema Servo No. 900-B volume ventilators were compared. The MA-1s had an average of 15,555 hours use prior to the study, compared with 200 average hours on each BEAR and 2,600 on each Servo ventilator. The MA-1s required the least number of repairs and operated longest between repairs, with 18,999 average hours of use per repair, compared with 4,955 hours for the BEAR-1 and 599 hours for the Servo ventilator. The average time required for repair was highest for the MA-1. Most repairs to the BEAR-1 were compressor-related and most were performed in the hospital, as were the repairs to the Servo ventilators. All MA-1 repairs were performed at Puritan-Bennett. The technical advantages of the BEAR-1 and Servo ventilators are convenient, but because of the frequency of repair of these units they are probably more suited to institutions where close supervision and technical backup are available.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Respiratory Volumes and Capacities