An evaluation of the P.K. Morgan pocket spirometer
Respiratory Care
|August 8, 1986
Summary
The P.K. Morgan pocket spirometer offers adequate accuracy for bedside screening of forced vital capacity (FVC) and forced expiratory volume in 1 second (FEV1). However, it is not accurate enough for diagnostic spirometry.
Area of Science:
- Pulmonary Function Testing
- Medical Device Evaluation
- Respiratory Diagnostics
Background:
- The P.K. Morgan pocket spirometer is a new, portable, and affordable device for measuring lung function.
- It measures key respiratory parameters: forced vital capacity (FVC) and forced expiratory volume in 1 second (FEV1).
Purpose of the Study:
- To assess the accuracy of FVC and FEV1 measurements obtained using the P.K. Morgan spirometer.
- To determine if the device meets established accuracy standards for pulmonary function testing.
Main Methods:
- The study utilized a Vent-Aid TTL Training Test Lung to generate airflow.
- Airflow was directed through the Morgan spirometer and a 9-L Collins water-sealed spirometer for comparison.
- Measurements were compared against a gold-standard spirometer to evaluate accuracy.
Main Results:
- Morgan spirometer FVC measurements showed high correlation (0.996) with Collins spirometer, with no significant difference (P=0.17).
- 56% of Morgan FVC measurements met American Thoracic Society (ATS) accuracy standards.
- Morgan spirometer FEV1 measurements also showed high correlation (0.999) with Collins spirometer, with a statistically significant but clinically small difference (P=0.01).
- 87% of Morgan FEV1 measurements met ATS accuracy standards.
Conclusions:
- The P.K. Morgan spirometer demonstrates adequate accuracy for bedside screening purposes.
- The device's accuracy is insufficient for use as a replacement for more precise pulmonary function equipment in diagnostic settings.