Accuracy of radial artery blood pressure determination with the Vasotrac
K G Belani1, J J Buckley, M O Poliac
1Department of Anesthesiology, University of Minnesota Medical School, Minneapolis 55455, USA. belan001@tc.umn.edu
Purpose:
To evaluate the accuracy of a new non-invasive method (Vasotrac) to measure blood pressure (BP) with accompanying arterial wave-form and pulse-rate display when compared with BP and waveform measured invasively.
Methods:
Healthy volunteers (n=53) served as subjects for the study. Blood pressures and waveforms obtained via a radial artery catheter (IABP) were compared with non-invasive measurements obtained every 12-15 beats by the new system (Vasotrac BP) from the opposite radial artery. In a sub-group of volunteers (n=11), BP was acutely increased and decreased with isoproterenol, phenylephrine and sodium nitroprusside. Data were analyzed by determining correlation and agreement between the two systems of measurement. Waveforms obtained by the two systems were qualitatively examined.
Results:
Non-invasive BP measured every 12-15 beats by the Vasotrac correlated with IABP (systolic r2 = 0.89; diastolic r2 = 0.88; mean r2 = 0.94). The actual values obtained by the two methods agreed closely with > 90% of readings being within 2SDs when plotted by the Bland Altman method. This was also true during vasoactive drug infusion when BP changed acutely and swiftly. During this dynamic period, Vasotrac BP accurately tracked the changes in IABP with correlations (systolic r2 = 0.82; diastolic r2 = 0.89; mean r2 = 0.95) and close agreement (> 90% of readings were within 2 SDs in the Bland Altman plot). Waveforms displayed by the two systems were qualitatively very similar. Pulse rates measured by the two systems were identical.
Conclusions:
The Vasotrac system displayed an arterial waveform which was similar to that obtained directly and measured BP and pulse rate accurately. It should be a convenient device to measure BP continually in a non-invasive fashion.
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure