Storage phosphor and film-screen mammography: performance with different mammographic techniques
S Kheddache1, A Thilander-Klang, B Lanhede
1Department of Diagnostic Radiology, Göteborg University, Sahlgrenska University Hospital, Sweden.
Abstract:
The aim of this study was to compare the image quality of storage phosphor plates with that in screen-film radiograms in mammography. Two anode/filter combinations were also compared--Mo/Mo and W/Rh. S Storage phosphor plates, generation IIIN (Fuji, Tokyo, Japan) and a conventional screen-film system (Kodak, Rochester, N. Y.) were evaluated using two mammographic units. One unit had a 0.6-mm focal spot, an anode/filter combination of Mo/Mo and no grid (AMo); the other had a 0.3-mm focal spot, a grid, and two possible combinations of anode/filter Mo/Mo (BMo) and W/Rh (BW). Simulated tumours and microcalcifications were randomly positioned in an anthropomorphic breast phantom (RMI model 165, no. 210-009, Radiation Measurements Inc., Middleton, Wisconsin). The image quality was evaluated using a modified version of receiver operating characteristics analysis. Five observers evaluated 300 films and 300 hard copy images each. Radiation doses were also determined. The image quality of the conventional screen-film images was significantly better than that for the storage phosphor plate mammograms. The BMo system rated best, for the detection of both tumours and microcalcifications, although it was not significantly different from the BW system. Systems BMo and BW rated significantly better than the AMo system for both image receptors studied. The mean absorbed dose was twice as high for the BMo system as for the AMo and BW systems for both conventional and digital technique. The mammograms produced with the screen-film combination gave a significantly better detectability than the storage phosphor plates used in this study. Substantial dose reduction could be achieved using an anode/filter combination of W/Rh instead of Mo/Mo with no significant loss of information in the images.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging