Related Experiment Video
Updated: Aug 14, 2026

Methods for the Discovery of Novel Compounds Modulating a Gamma-Aminobutyric Acid Receptor Type A Neurotransmission
Published on: August 16, 2018
Automated microanalysis of gabapentin in human serum by high-performance liquid chromatography with fluorometric
P H Tang1, M V Miles, T A Glauser
1Department of Pediatric Neurology, The Children's Hospital Medical Center, Cincinnati, OH 45229-3039, USA. Tangp0@chmcc.org
Abstract:
An automated high-performance liquid chromatographic method for the determination of gabapentin, 1-(amino-methyl)cyclohexaneacetic acid, in serum is described. The procedure involves protein precipitation with methanol followed by using a robotized derivatization with o-phthaldialdehyde reagent and automated high-performance liquid chromatography. The analog of gabapentin, 1-(aminomethyl)cycloheptaneacetic acid, was used as the internal standard. Blank serum was fortified with gabapentin (0.1-10.0 microg/ml) and internal standard. Separation was achieved on a Waters 5-microm reversed-phase column (10 cmx4.6 mm) with mobile phase consisting of 0.02 M phosphate buffer (pH 4.5)-acetonitrile (50:50, v/v). Eluents were monitored by fluorescence spectroscopy with excitation and emission wavelengths of 230 and 420 nm, respectively. The calibration curve for gabapentin in serum was linear (r=0.999) over the concentration range 0.1-10.0 microg/ml. The inter- and intraassay variations for three different gabapentin concentrations were < or =10% throughout. The lower limit of quantitation was found to be 0.1 microg/ml. Chromatography was unaffected by a range of commonly employed antiepileptic drugs or selected amino acids.