An unexpected complicationof the remote-controlled intraurethral valve pump for urinary incontinence
1Urologische Klinik der Ruhr-Universität Bochum, Marienhospital Herne, Deutschland.
Abstract:
We report an unexpected complication of the remote-controlled intraurethral valve pump in a patient with neurogenic bladder. A meningomyelocele patient with an atonic bladder received an intraurethral valve pump. Acute urinary retention was caused by a mucus clot obstructing the pump. Prior to extended clinical use, the remote-controlled intraurethral valve pump should be evaluated in prospective multicenter studies.
Related Concept Videos
The Micturition Reflex
The process begins with bladder filling, where the bladder wall stretches as urine accumulates. This stretching activates the urine storage reflex, mediated by the sacral spinal segments and the pontine storage center. Efferent sympathetic impulses stimulate the detrusor muscle to relax and the internal urethral sphincter to contract, facilitating urine...