On-line technique for measuring stable oxygen and hydrogen isotopes from microliter quantities of water
R A Socki1, C S Romanek, E K Gibson
1Lockheed Martin, Houston, Texas 77058, USA.
Abstract:
Detailed here is a method for extracting and analyzing oxygen and hydrogen isotopes from 10 microL-sized water samples. Based on the traditional CO2-H2O equilibration technique, the oxygen isotope exchange reaction is done exclusively in sealed 6-mm (o.d.) Pyrex tubes at 25 degrees C, with full isotope exchange completed in at least 28 h. Using the same water sample employed in the 18O equilibration, D/H extractions are done in separate sealed 6-mm (o.d.) Pyrex tubes by reaction with Zn at 450 degrees C to form H2(g). Provided that a correction factor is applied to 18O analyses, accuracy and precision for both 18O and D/H are comparable to standard techniques using much larger samples.