Preliminary tests and construction of a computerized quantitative surgical keratometer
Related Experiment Video
Updated: Oct 20, 2026

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
L Carvalho1, S A Tonissi, J C Castro
1Instituto de Física de São Carlos-USP, Brazil.
Purpose:
To construct a simple, computer-based, quantitative surgical keratometer to measure a 3.0 to 4.0 mm central region of the corneal surface.
Setting:
Laboratorio de Optica Oftalmica, Instituto de Física de São Carlos (IFSC-USP), São Paulo, Brazil.
Methods:
A high-intensity fiber-optic-illuminated ring pattern (Placido disk) is projected on the cornea. Reflected images are captured by a charge-coupled device camera mounted on a Zeiss microscope beam splitter and then digitized by a frame grabber installed on an IBM-compatible personal computer. Simple algorithms based on image-processing techniques were implemented for border detection. A calibrating curve based on 4 spherical surfaces was used to calculate diopter values at 360 points at each examination. Results were plotted on the computer monitor using diopter value versus angle (1 through 360 degrees) graphs. Preliminary measurements of 14 healthy corneas were compared with the equivalent radial distance points measured on an EyeSys corneal topographer.
Results:
Mean deviation was 0.05 mm for radius of curvature, 0.24 diopter for power, and 5 degrees for cylinder.
Conclusion:
The keratometer provided precise measurements for corneal shape control during surgery.