The midline dose distribution for a three-field radiotherapy technique
S L Meeks1, R O Williams, F J Bova
1Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, USA.
Summary
The University of Florida
Area of Science:
- Radiation oncology
- Medical physics
- Head and neck cancer treatment
Background:
- Head and neck cancer radiotherapy commonly uses three fields.
- A potential risk is radiation myelitis due to field overlap.
- The University of Florida employs a specific three-field technique.
Purpose of the Study:
- To evaluate radiation dose at the junction of three treatment fields.
- To assess the risk of radiation myelitis in head and neck cancer patients.
- To validate the safety of the University of Florida's radiotherapy technique.
Main Methods:
- Irradiation of head and neck phantoms using University of Florida setups.
- Utilizing anthropomorphic phantoms and tissue-equivalent materials.
- Employing a two-dimensional thermoluminescent dosimeter array for dose measurement.
Main Results:
- No excess radiation dose was detected at the junction of the three fields.
- Spinal cord dose ranged from 10% to 100% of the primary field's central axis dose.
- Specific dose percentages were recorded for oropharynx and larynx setups.
Conclusions:
- The University of Florida's three-field technique does not increase radiation dose at field junctions.
- This technique poses no measurable additional risk of radiation myelitis.