[The current state of leak in anesthetic machines detected by low flow leak tests]
1Department of Anesthesiology, Faculty of Medicine, University of the Ryukyus, Okinawa.
Abstract:
To assess the current state of leak in anesthetic machines, we selected 66 units of anesthetic machines for inspection and repair from various medical institutions. Based on a newly designed inspection flow chart a low flow leak test for internal circuits of the anesthetic machines was performed. The conventional low flow leak test was also performed for smooth detection of leak for rational evaluation. Only 39% of the anesthetic machines met the standard of the low flow leak tests, and leak was detected in the remaining 61%. The average residual leak mounted to 0.97 l.min-1, with the maximum of 5.3 l.min-1. Canisters, corrugated tubes, and vaporizers were considered the primary causes of leak. After the inspection and repair, leak in 77.5% of the anesthetic machines either disappeared or decreased and the average residual leak dropped to 0.34 l.min-1. However, 47% of the anesthetic machines still failed to meet the standard of the low flow leak tests. To further improve the situation, more detailed inspection and repair are necessary especially for precise detection of the cause of leak in the internal circuit of anesthetic machines which often remains undetected.