Related Experiment Video
Updated: Aug 13, 2026

12:20
Use of the Operant Orofacial Pain Assessment Device (OPAD) to Measure Changes in Nociceptive Behavior
Published on: June 10, 2013
An algometer for intraoral pain tolerance measurements
C H Gibbs1, K Karpinia, J E Moorhead
1Department of Oral Biology, College of Dentistry, University of Florida, Gainesville 32610, USA. cgibbs@dental.ufl.edu
Journal of Neuroscience Methods
|July 2, 1999
Summary
A new intraoral algometer precisely measures pain tolerance (PT) in oral tissues. Reproducibility improved with subject familiarity, showing anterior sites are more sensitive than posterior ones.
Area of Science:
- Biomedical Engineering
- Neuroscience
- Oral Medicine
Background:
- Accurate measurement of intraoral pain perception is crucial for understanding oral conditions.
- Existing methods may lack precision or intraoral applicability.
- Development of a specialized device is needed for reliable sensory testing in the mouth.
Purpose of the Study:
- To develop and validate a novel intraoral algometer for measuring pain tolerance (PT).
- To assess the reproducibility of PT measurements in healthy subjects.
- To compare pain sensitivity across different intraoral sites.
Main Methods:
- An intraoral algometer with a 0-200 g force range and a 0.40 mm hemispherical tip was designed.
- Nine healthy adults underwent 6 weekly sessions measuring PT at eight intraoral sites.
- Reproducibility was assessed using interclass correlation coefficients (R), and sensitivity differences were analyzed using ANOVA and Duncan's test.
Main Results:
- Reproducibility (R values) improved from 0.40-0.79 (all visits) to 0.63-0.97 (experimental visits 3-6).
- No significant interaction was found between measurement site and visit.
- Anterior intraoral sites demonstrated significantly higher sensitivity (P < 0.05) compared to four posterior sites.
Conclusions:
- The developed intraoral algometer offers precision, computerized data entry, and an autoclavable handpiece, making it suitable for oral investigations.
- PT measurements showed good to excellent reproducibility after a calibration period.
- The device shows potential for broader sensory and pain research beyond oral applications.