Broad-band spectral analysis of 24 h continuous finger blood pressure: comparison with intra-arterial recordings
Related Experiment Video
Updated: Oct 19, 2026

Long-term Blood Pressure Measurement in Freely Moving Mice Using Telemetry
Published on: May 17, 2016
P Castiglioni1, G Parati, S Omboni
1LaRC, Centro di Bioingegneria, Fondazione Don C. Gnocchi and Politecnico di Milano, via Gozzadini 7, I-20148 Milan, Italy.
Abstract:
The present study compares the spectral characteristics of 24-h blood pressure variability estimated invasively at the brachial artery level with those estimated by measurement of blood pressure at the finger artery using the non-invasive Portapres device. Broad-band spectra (from 3x10(-5) to 0.5 Hz) were derived from both finger and intra-brachial pressures recorded simultaneously for 24 h in eight normotensive and twelve hypertensive ambulant subjects. At frequencies lower than 0.07 Hz, higher spectral estimates were obtained by Portapres than by intra-brachial measurements. The maximum overestimation occurred in systolic pressure at around 10(-2) Hz, where the amplitude of the oscillations was two times greater when measured by Portapres. A less pronounced overestimation was found for diastolic pressures. The maximum overestimation was greater during daytime than during night-time. At around 0.1 Hz, invasive and non-invasive spectra were similar. At the respiratory frequencies (0.15-0.50 Hz), the power spectra were overestimated by Portapres during daytime, and underestimated at night. These results provide reference information for the correct interpretation of Portapres data in the estimation of 24-h blood pressure spectral power.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...