Effect of surface coating on platelet count drop during cardiopulmonary bypass
Related Experiment Video
Updated: Oct 19, 2026

Artificial Lung Device Priming for In Situ Fiber Bundle Surface Grafting
Published on: March 28, 2025
D A Palanzo1, D L Zarro, N J Manley
1Department of Perfusion, Lehigh Valley Hospital, Allentown, Pennsylvania 18105, USA. David.Palanzo@LVH.com
Abstract:
This study was designed to investigate the effect of surface coating on platelet count drop during cardiopulmonary bypass (CPB). Sixty patients undergoing open-heart surgery were randomly divided into three groups each receiving a different type of coated hollow-fiber membrane oxygenator. The patients were given either an uncoated oxygenator (noncoated group), an oxygenator coated with Carmeda (Carmeda group) or an uncoated oxygenator with albumin in the priming solution (albumin group). Comparisons were made in platelet count pre-CPB, on bypass (15-25 min) and during the warming period. Calculations were used to account for the effect of hemodilution. The albumin group had significantly lower platelet count drops (-4.8+/-7.1%) than the Carmeda group (11.0+/-8.3%) and the noncoated group (20.3+/-14.5%). Carmeda surface coating demonstrated some beneficial effects, but to a lesser degree than the albumin.