Background:
Non-invasive oxygen monitoring with pulse oximetry or transcutaneous monitoring has gained widespread use in neonatology. Different factors like arterial hypotension, peripheral vasoconstriction and edema adversely affect the accuracy of both methods. To ensure reliable monitoring of oxygen saturation in critically ill patients we measured oxygen saturation with a fiberoptic catheter via umbilical artery.
Methods:
In ventilated premature infants (FiO2 > 0.4) a 4F-fiberoptic catheter (Oximetrix)-3, Abbott) was inserted to the descending aorta (Th 6-8). Simultaneously pulse oximetry (SaPO2) was performed with the Ohmeda Biox 3700. To compare the reliability of both methods, blood was analysed for arterial partial oxygen pressure (PaO2), fetal hemoglobin (HbF) and arterial oxygen saturation (SaO2) by complete co-oximetry (Radiometer Copenhagen OSM3) as reference.
Results:
In 10 premature infants (median gestational age 30.5 weeks; median birth weight 1360 g) oxygen saturation was measured with the fiberoptic catheter (SaFO2) over a total period of 935 hours. In all, 137 blood samples were analysed for arterial saturation (SaO2) by co-oximetry. The mean difference between the SaO2 and SaFO2 was -1.89% (+/- 1.53); the mean difference between SaO2 and the values obtained by pulse oximetry (SaPO2) was -3.09% (+/- 2.33). The SaFO2 results correlated closely with the co-oximetry values (r = 0.97; p < 0.0001).
Conclusion:
In critically ill patients, if non-invasive oxygen monitoring fails, a fiberoptic catheter offers the possibility of continuous and reliable measurement of oxygen saturation.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...