Piroxicam quantitation in human plasma by high-performance liquid chromatography with on- and off-line solid-phase extraction
Related Experiment Video
Updated: Oct 19, 2026

Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
M Yritia1, P Parra, J M Fernández
1Analysis Laboratory, Sant Pau Hospital, Barcelona, Spain.
Abstract:
A comparative study of two analytical methodologies for piroxicam quantitation in plasma by off-line and on-line solid-phase extraction (SPE) and high-performance liquid chromatography (HPLC) is described. The SPE cartridges contained C8 for both extraction methods. The analytes piroxicam and tenoxican (internal standard) were separated on a C18 column with a mobile phase consisting of acetonitrile:20 mM phosphate buffer pH 3.1 (50:50, v/v) followed by UV detection at 360 nm. The validation of the methods demonstrated good recoveries (over 90%), sensitivity (limits of quantification of 0.05 microgram/ml with on-line SPE and 0.1 microgram/ml with off-line SPE, based on a 100 microliters and 200 microliters sample volume, respectively), accuracy and precision (better than 9.5%). Both methodologies have been used for bioequivalence studies.