Planning target volumes for radiotherapy: how much margin is needed?
1Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston 77030, USA.
Summary
This study determined the ideal planning target volume (PTV) margin for radiotherapy by considering only physical uncertainties. A margin of 1.65 standard deviations (SD) ensures the clinical target volume (CTV) is adequately covered, optimizing tumor treatment and sparing healthy tissues.
Area of Science:
- Radiotherapy physics and planning
- Medical imaging and uncertainty quantification
- Radiation oncology
Background:
- Radiotherapy planning target volume (PTV) margins account for uncertainties like patient positioning and organ motion.
- Current PTV margins often consider doses to adjacent healthy tissues, deviating from ideal physical uncertainty-based margins.
- This study focuses on calculating PTV margins based solely on physical position uncertainties.
Purpose of the Study:
- To calculate the ideal CTV-PTV margin size considering only physical position uncertainties.
- To evaluate different strategies for determining CTV-PTV margins based on probability.
- To provide a data-driven recommendation for PTV margin selection in radiotherapy planning.
Main Methods:
- Modeled clinical target volume (CTV) position using independent Gaussian distributions in three Cartesian directions.
- Analyzed three margin strategies: complete CTV enclosure, beam's eye view (BEV) projection enclosure, and edge point inclusion.
- Derived cumulative probability distributions for each margin strategy.
Main Results:
- A 1 SD margin resulted in 24% CTV enclosure, 39% BEV projection enclosure, and 84% edge point inclusion.
- A 95% probability of complete CTV enclosure requires a 2.8 SD margin.
- A 95% probability for BEV projection enclosure requires 2.45 SD; for edge point inclusion, it requires 1.65 SD.
Conclusions:
- The third strategy (edge point inclusion) correlates margin size with minimum target dose.
- A margin of 1.65 SD is recommended for calculating PTV from CTV, ensuring >95% minimum CTV dose.
- Modifying beam portals, not PTV margins, should be used for sparing healthy structures, improving dose distribution accuracy.