Levels of naturally occurring microorganisms on surgical instruments after clinical use and after washing
N S Chu1, H Chan-Myers, N Ghazanfari
1Advanced Sterilization Products, Irvine, CA 92618-9824, USA.
Abstract:
Surgical instruments exposed to sterile body sites should be contaminated with relatively low levels of microbial contamination or bioburden; however, few studies in the literature have determined the quantitative level and types of contamination. A study was conducted at 2 clinical sites to determine the level of microbial contamination of surgical instruments after clinical use and after washing. Quantitative assays showed that bioburden levels were in the range of 0 to 4415 colony forming units per instrument after clinical use, and 88% of the instruments had bioburden levels lower than 1000. As expected, a reduction in counts occurred after washing; however, in some cases, higher counts were found on the instruments after the washing process. Although the washing procedure is effective in reducing the microbial levels deposited on the surgical instruments during use, a recontamination process occurs that results in increased counts after washing. The low bioburden level after washing consists of predominantly vegetative microorganisms that present a relatively low challenge to sterilization and disinfection systems.
More Related Videos
Related Concept Videos
Asepsis
Hand hygiene
Hand washing...
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated.
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Handwashing III: During the Procedure and Post-Procedure Steps