Clinically shaped fields: relative outputs for Varian multileaf collimators and tray-mounted cerrobend blocks
K H Meyer1, S E Skubic, L C Baird
1Radiological Associates of Sacramento Medical Group Incorporated, Radiation Oncology Center, California 95816, USA. klaudia_meyer@hotmail.com
This study compared radiation output factors from multileaf collimators (MLC) and cerrobend blocks, finding minimal variation. MLCs offer a consistent alternative to blocks for shaping radiation fields.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Physics
Background:
- Multileaf collimators (MLCs) and cerrobend blocks are used to shape radiation beams in radiotherapy.
- Variations in relative output factors between these two methods can impact dose delivery accuracy.
Purpose of the Study:
- To quantify the variation in relative output factors between MLC-shaped fields and cerrobend block-shaped fields.
- To compare the dosimetric consistency of MLCs versus traditional blocks in clinical scenarios.
Main Methods:
- Compared 21 clinical fields from 16 patients using Varian linear accelerators (2300CD and 2100CD).
- Measured output factors in solid water at specified depths and source-to-axis distance (SAD = 100 cm) for 6 and 20 MV photon beams.
- Utilized cerrobend blocks cast from films that generated the MLC shapes for direct comparison.
Main Results:
- The overall average MLC factor (MLC reading/block reading) was 0.9952 with a standard deviation of 0.0024.
- Clinical average MLC factors were 0.9957 (SD=0.0022), consistent across different energies and depths.
- Repeat measurements on a different MLC model showed results within 0.2% of initial findings.
Conclusions:
- MLC-shaped fields demonstrate high consistency and minimal variation compared to cerrobend blocks.
- MLCs provide a reliable and accurate method for field shaping in radiotherapy, comparable to traditional blocks.
- Further research may explore correlations between MLC factors and open field percentages.