Related Experiment Video
Updated: Jul 29, 2026

Endoscopic Bilateral Nipple-sparing Mastectomy via a Single Axillary Incision with Immediate Pre-pectoral Implant-based Breast Reconstruction
Published on: May 17, 2024
Buckled upper pole breast style 410 implant presenting as a manifestation of capsular contraction
1The Cosmetic and Restorative Surgery Clinic, NSW, Sydney, Australia.
Abstract:
Recent experience with the cohesive gel implant has shown an interesting variation of capsular contracture. Instead of spherical contraction, the capsule formation around the 410 implant seems to cause a buckling of the superior pole of the implant. This buckling has been observed in four implants and presented clinically in 2 of 50 patients with a superior pole mass. The capsular contraction around the implant and the buckling are probably camouflaged by a subpectoral implantation and this presentation should be recognized by the surgeon, especially when considering placement of the cohesive gel implant in a subglandular position. Histological sections of a capsule around the implant show fragmented silicone, with a cellular reaction around the fragmented silicone.
More Related Videos
07:10Modified Long Head of Biceps Tendon Rerouting and Fixation as Partial Capsular Reconstruction for Massive Irreparable Rotator Cuff Tears
Published on: March 6, 2026
03:07Single-Port Robotic-assisted Transaxillary Breast-conserving Surgery: A Prospective, Single-arm, Non-randomized Phase IIa Clinical Trial
Published on: August 19, 2025