Comparison of interface relationships between implant components for laser-welded titanium frameworks and standard cast frameworks
Related Experiment Video
Updated: Aug 16, 2026

Multi-Scale Modification of Metallic Implants With Pore Gradients, Polyelectrolytes and Their Indirect Monitoring In vivo
Published on: July 1, 2013
1Department of Prosthodontics, University of Washington, School of Dentistry, Seattle 98195, USA.
Abstract:
With the introduction of new techniques for the fabrication of frameworks for implant-supported prostheses comes the need to understand how the components used compare to those used for conventional cast frameworks. The relationship of measured machining tolerances between conventional implant components and those components used for stereo laser-welded implant frameworks was determined using a standardized protocol. Statistically significant differences in the horizontal interface relationship were found between paired implant components, which had a mean range from 23.1 to 51.7 microns. From a laboratory and clinical perspective, machining tolerances of implant components represent a variable associated with their manufacturing, which can ultimately affect the fit of a completed prosthesis.