Related Experiment Video
Updated: Aug 5, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Patellar contact forces with and without patellar resurfacing in total knee arthroplasty
R Singerman1, S M Gabriel, C B Maheshwer
1Department of Orthopaedics, Case Western Reserve University, Cleveland, Ohio 44106, USA.
Abstract:
Patellar contact forces were measured in 7 fresh-frozen cadaver knees during continuous flexion-extension cycling for the normal knee and after total knee arthroplasty without and with patellar resurfacing using 2 prosthesis systems with different designs for the femoral and patellar components. Analysis of variance with specimen as a repeated measure was used to examine the effect of contact surface. When compared to normal knees, total knee arthroplasty without patellar resurfacing resulted in no change in patellar contact forces. The magnitude of the inferosuperior patellar shear force increased significantly after patellar replacement with an oval, biconcave patellar component and increased further after replacement with an oval, domed patellar component. There were no significant changes in the mediolateral and anteroposterior contact force components after patellar replacement.